Warming large media bottles in Bath Armor.

 

Modified: July 10, 2008

Article: BT1426

 

Symptoms

Cell Culture media bottles take longer to bring to temperature in Bath Armor than in a standard water bath. In direct comparison to water, it takes Bath Armor approximately 1.25X longer to warm the contents of microfuge tubes and up to 2-3x longer to warm a 500 ml media bottle. To troubleshoot these issues, follow the recommendations below.

 

Problem

Due to its greater thermal mass, Bath Armor takes longer to warm than water. When a cold vessel is placed into Bath Armor, like water, it cools slightly then returns to temperature as it brings the vessel to temperature. Both the temperature drop and the rate of return to temperature depend on several factors related to (1) the volume and the original temperature of the vessel, (2) the amount of Bath Armor contained within the bath, (3) the set instrument temperature, and (4) the thermal efficiency of the bath.

 

Products Affected: Bath Armor

 
 
Resolution
Adjust the experimental protocol or increase the temperature of the bath.

To warm refrigerated vessels such as 500 ml tissue culture media bottles more quickly, simply raise the temperature of Bath Armor™ to compensate for the slower rate of heat transfer. For example, to raise the temperature of a typical 500 ml bottle of cell culture media from 4șC to approximately 37șC in 30 minutes, submerge the bottle in 50-60șC Bath Armor for the 30 minutes, then pull out and place the bottle onto the surface of Bath Armor. The Bottle of media will remain at 37 șC +/- 2șC atop the 50-60șC Bath Armor until use. But, to ensure proper heating of the sample and to avoid overtemperature exposure, be sure to validate the adjusted protocol and to keep the conditions constant from experiment to experiment.

 
To thaw frozen 500 ml bottles such as serum bottles more quickly, first bring the bottle to 4șC in a refrigerator prior to adding to the bath. This can effectively reduce the amount of time it takes to warm the bottle of serum from 1+ hours to 30-40 min. Additionally, by periodically relocating a cold bottle within the Bath Armor, the bath is able work more efficiently, which can reduce warm up times even more. It is best practice to warm larger media (500 ml) bottles in higher capacity baths. It takes longer to warm when the size ratio of the bath to the media bottle is too small. In some cases, small water baths, whether water-filled or waterless, can become very inefficient when multiple large bottles are thawed or warmed at the same time. Dual tank baths are ideal for laboratories that perform frequent thawing (see below).
 
 
Use a dual tank bath.
 
In some laboratories, a dual tank bath is the best option. Both tanks are filled with Bath Armor but set to different temperatures. One tank is set to 50șC for thawing and rapid warming while the other is set to the appropriate incubation temperature (37șC). Once vessels are thawed they can be transfered from the first tank to the second tank and allowed to rest until use.
 
 
For more information on troubleshooting these issues, email our technical services using the by clicking here Technical Assistance.
 
 
 
SupportIcon
Store Store-Nav Store Link2rollover2 Who's Armored Customers-Nav Who's Armored Link2rollover2b Free Test Drive TestDrive-Nav Bead Bath Link2rollover2b1 Beads Beads-Nav Lab Care Link2rollover2b1a Water Bath Home-Nav Water Bath Link2rollover2b1d Trays + Buckets Tray-Nav Personal Gear Link2rollover2b1b Lab Coats LabGear-Nav Support Link2rollover2b1c Support Support-Nav